Comparison of Pull-out Strength for Different Bone Block Length in a Porcine Anterior Cruciate Ligament Model

MAJ Matthew Posner [et al.] - Nama Orang;

Background: Bone block length for bone–patellar tendon–bone (BPTB) anterior cruciate ligament (ACL) reconstruction has traditionally been 25mm in length. The previous surgical technique did not require the surgeon to pay particular attention to the length of the bone block, and therefore, there is scant evidence in the literature describing ideal lengths. With the gaining popularity of accessory medial portal drilling of the femoral tunnel, concerns with tunnel length and graft shuttling have surfaced. Newer techniques have advised shortening of the femoral bone block to accommodate the shorter tunnel and for ease of bone block manipulation into the aperture of the tunnel.
Purpose: To compare the effects of bone block length on the pull-out strength of patellar tendon grafts using metal interference screws in a porcine ACL reconstruction model. The hypothesis was that the pull-out strength of each length of bone block under cyclic and ultimate load to failure testing would surpass the physiologic loads experienced by a normal ACL.
Study Design: Controlled laboratory study.
Methods: This study used 27 unmatched porcine femurs and BPTB constructs. Specimens were randomly assigned to a 10-, 15-,or 20-mm bone block reconstruction and a cycle load of 100, 500, or 1000 cycles. This resulted in 9 specimen groups with 3 specimens in each group. A central composite design (CCD) for the test matrix was selected, as this was optimum for requiring relatively few experiments while still exploring the complete range of interest for 2 independent variables. Each reconstruction used a 7 20–mm titanium interference screw. All reconstructions were performed on the femoral side using 10-mm-wide patellar tendon grafts, and tensile tests were performed. The loading protocol started with a 20-N preload, then cyclic testing to the appropriate number of cycles in the elastic region between 50 and 150 N at a strain rate of 200 mm/min, and then ended with ultimate load-tofailure
testing. Ultimate load to failure, peak stress, elongation, and stiffness were all recorded. The patellar tendon graft mode of failure was measured by visual inspection.
Results: During load-to-failure testing, 5 of 9 graft constructs in the 10-mmgroup failed at the bone block, while 2 of 9 failed in the 15-mm group at this interface. In the 20-mm group, all 9 specimens failed at the tendon, and none failed at the bone block. There was a statistically significant difference in modes of failure between the bone block length groups in the reconstructed ACL grafts. Analysis
indicates that a smaller bone block length graft ismore likely to fail due to a bone block failure than a tendon failure. The average ± standard deviation failure load for all specimenswas 573 ± 171 N.Themean failure loads for the 10-, 15-, and 20-mmgroupswere 614 ± 110,658 ± 92, and 540 ± 203 N, respectively. There was no statistical significance between the groups in any of these measurements.
Conclusion: Bone blocks of 20 to 25mmin length are normally used in surgical practice. Thus, the purpose of this studywas to explore the effects of asmaller boneblock length in the fixation strength of a graft. This study could not yielda significant difference in failure load for differently sized bone blocks. There was a significant tendency of shorter bone block lengths to fail due to bone block failure. Clinical Relevance: Bone block failure was defined as slippage of the bone block or interference screw. These results show that using a smaller bone block may increase the likelihood of a graft failure in an ACL reconstruction.

Keywords: ACL reconstruction; BTB autograft; interference screw fixation


Ketersediaan

Tidak ada salinan data

Informasi Detail
Judul Seri
The Orthopaedic Journal of Sports Medicine
No. Panggil
-
Penerbit
: The Author., 2014
Deskripsi Fisik
The Orthopaedic Journal of Sports Medicine, 2(5), 2325967114532762
Bahasa
English
ISBN/ISSN
DOI: 10.1177/2325967
Klasifikasi
-
Tipe Isi
-
Tipe Media
-
Tipe Pembawa
-
Edisi
-
Subjek
KESEHATAN
Info Detail Spesifik
-
Pernyataan Tanggungjawab
Wati/Agus
Versi lain/terkait

Tidak tersedia versi lain

Lampiran Berkas
  • FULL TEXT:Comparison of Pull-out Strength for Different Bone Block Length in a Porcine Anterior Cruciate Ligament Model
Komentar

Anda harus masuk sebelum memberikan komentar

Elibrary Perpustakaan Universitas Riau
  • Informasi
  • Layanan
  • Pustakawan
  • Area Anggota

Tentang Kami

As a complete Library Management System, SLiMS (Senayan Library Management System) has many features that will help libraries and librarians to do their job easily and quickly. Follow this link to show some features provided by SLiMS.

Cari

masukkan satu atau lebih kata kunci dari judul, pengarang, atau subjek

Donasi untuk SLiMS Kontribusi untuk SLiMS?

© 2025 — Senayan Developer Community

Ditenagai oleh SLiMS
Pilih subjek yang menarik bagi Anda
  • Karya Umum
  • Filsafat
  • Agama
  • Ilmu-ilmu Sosial
  • Bahasa
  • Ilmu-ilmu Murni
  • Ilmu-ilmu Terapan
  • Kesenian, Hiburan, dan Olahraga
  • Kesusastraan
  • Geografi dan Sejarah
Icons made by Freepik from www.flaticon.com
Pencarian Spesifik
Kemana ingin Anda bagikan?